GE Healthcare · CT183177230000065
Scanner Hardware Stopped Scan
Symptom
Scanning aborted with Buffer overrun detected during scan data transfer
Error message
Error: 183177 Abnormal Xray termination. External x-ray interlock opened See the long description for the full error code Tue Sep 6 09:47:22 2005 Error: 183177 Host: OBC Process: Generator File: gen_xray.c 1.31 Line: 604 Function: XRAY GENERATION : Xray Exposure Interlocks Scan: 637/7/1 Type: Helical Exception Level: Pri/Most Severe Time: 09:47:22:260 Log Series: 487 Abnormal Xray termination. External x-ray interlock opened. Error: 230000065 Buffer overrun detected during scan data transfer. See long description for full error Error: 230000065 Buffer overrun detected during scan data transfer. This is generally not a hardware issue and will not be remedied by replacing or otherwise servicing the gantry, fiber-optic cable or DIP card. Interrupt bit: 6 Violation count: 1 Scan aborted.
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2225004-9 73GB scan disk (replaces 2225004-8 and 18 GB p/n 2225004-4 and 36 GB scan disk 2225004-5) See Long Description for Details. Part #: 2225004-9 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73GB 15000 RPM Scan Data Drive FRU: Yes Status: Orderable Comments: Per ECR 2046173, 212008898, 8/8/07. ** FE ASSIST ** Refer to Service Note SN07110612a For Jumper information (12/3/07 212001261) Part #: 2225004-8 Part Type: GP CONSUMABLES Part Weight: 1.61 LBR Repairable: No Returnable: No Hazardous Material: N Description: HARD DRIVE 73GB 15KRPM SCAN DATA FRU: Yes Status: Orderable Comments: Added per ECR 2031409, 7/14/06, ms11892. ** FE ASSIST ** If no stock of the 2225004-8 order 2225004-9 as the replacement. Refer to Service Note SN07110612a (212001261 12/3/07)
- Area
- CT
Full cause & step-by-step resolution