Pular para o conteúdo principal
GE Healthcare · MROptima22735512273550

Unable to scan CUBE with RF Hub Power Failure errors due to Defective Coil.Replace 8 Channel Head Coil,wqr

Symptom

Unable to scan CUBE with RF Hub Power Failure errors

Error message

2273551 RF Hub Power Failure. The RF Hub detected a power failure on the following board. RF Hub (board name), power +-%V.

2273550 RF Hub Power Failure see long Error: 2273550 Host: WBH_RAD_MR11 Process: NSP File: SCP:RfHubHw::Pdb Line: 734 Exception Class: Unknown Severity: Unknown Function: RF: RF Receive PSD: Unknown Coil: Unknown Scan: -1/-1 RF Hub Power Failure. The RF Hub RFCB read that the PDB is reporting power failures for the following power distributions. Common: OK RFCB: -10V MCDB: -10V RFSB: OK->

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5146634-2 1.5T HD 8CH HiRes Brain Coil with Baseplate and Mirrors 5146634-2 1.5T HD 8CH HiRes Brain Coil with Baseplate and Mirrors.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.