PURE Images Blank due to 1.5T Body Hybrid Splitter.
Symptom
PURE Images Blank See Long Description PURE images blank -- just a few shades of gray, no anatomy. Only PURE is affected. All images in the PURE series are affected. Applications reviewed the protocol and said it was okay. If the customer removes the PURE scan option, the images are normal. The errors shown below are logged. See attachment for an image from this site showing the problem. Thu Jun 23 19:12:54 2011 2270061 2 MR01 LxMxConsumer VrfTrace.cpp 93 Server Name: ADS Signal overrange detected (R1 analog) for channel 2 from Receiver ID 0. The RRX is connected to VRF located at ICN1(ICN2), if ADS(ADS1) reports this error. Thu Jun 23 19:12:54 2011 2270060 1 MR01 LxMxConsumer VrfTrace.cpp 113 Server Name: ADS MR Signal too high (overrange) detected.
Error message
2270061 Signal overrange detected (R1 analog) for channel 2 from Receiver ID 0. See Long Description 2270061 MR01 LxMxConsumer VrfTrace.cpp 93 Server Name: ADS Signal overrange detected (R1 analog) for channel 2 from Receiver ID 0. The RRX is connected to VRF located at ICN1(ICN2), if ADS(ADS1) reports this error. 2270060 MR Signal too high overrange detected. See Long Description 2270060 MR01 LxMxConsumer VrfTrace.cpp 113 Server Name: ADS MR Signal too high (overrange) detected.
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5342686 Hybrid Collector, Body, 1.5T 5342686 Hybrid Collector, Body, 1.5T
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.