GE Healthcare · CTBrivo186328
Scanner Hardware Stopped Scan with Servo Amp alarm
Symptom
Scanner Hardware Stopped Scan with Servo Amp alarm Scanner Hardware Stopped Scan with Servo Amp alarm. No further details Scanner Hardware Stopped Scan with Servo Amp alarm
Error message
Error: 186328 13-1002-04 Gantry Req Error: ??? Servo amp Alarm is detected See the Long Description for the complete error. Error: 186328 13-1002-04 Gantry Req Error: ??? Servo Amp alarm is detected Gantry Req Error: 4 or 7 Code: 13-1002-04 Name: Servo Amp Alarm. (GP) Class: Error Description: Servo Amp alarm is detected. Potential Cause / FRU 1. Too many acceleration/deceleration in short duration. 2. Servo Amp 3. Axial Motor
Applies to
- Models
- Brivo
- Product
- BRIVO CT315 / 325 See Long Description BRIVO CT315 / 325 List of softwares the product can run: V/R 1.48 V/R 1.50 V/R 1.53 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5323475 Axial drive regen resistor-use on Brivo CT315 / CT325 5323475Axial Drive Regen Resistor--Use on Brivo CT315/CT325 Part #: 5323475Part Type: GP CONSUMABLES Part Weight: 0.00 KGMRepairable: NoReturnable: NoHazardous Material: NDescription: Axial drive regen resistorFRU: YesStatus: Orderable
- Area
- CT