GE Healthcare · CT160472687225588

Lightspeed 1.X - Octane

Symptom

System posts: Edit Patient Data cannot remove the exam entry: it may be in use. System posts Edit Patient Error message: Edit Patient Data cannot remove the exam entry: it may be in use.

Error message

Error: 16047 [EPD] Failed to remove the exam entry See the LongDescription for the complete error code Error: 16047 [EPD] Failed to remove the exam entry Sat Nov 27 19:30:24 2004 Error: 16047 Host: ct04 Process: editPatientData File: Line: 0 [EPD] Failed to remove the exam entry

Error: 1031 stored procedure failed See Long Description for Details. Sun Nov 14 20:00:37 2004 Error: 1031 Host: mcas Process: dbrserver stored procedure failed : -284 - getexamuidinfo failed Also check usr/g/ctuser/logfiles/epdlog with command: tail -100 /usr/g/ctuser/logfiles/epdlog | more MESSAGE from Process 26872>> Sun Nov 14 20:12:15 2004 (netclient.c:93) [Client]> The request manager started MESSAGE from Process 26872>> Sun Nov 14 20:12:48 2004 (:0) [EPD] Failed to remove the exam entry Thu Jul 17 13:45:18 2008 Error: 1031 Host: ct99 Process: dbrserver File: batch.ec Line: 599 stored procedure failed : -243 - Fetch failed for image - ISAM code : -143 Wed Jul 2 10:54:38 2008 Error: 1031 Host: CT01 Process: dbwserver File: batch.ec Line: 934 stored procedure failed : -25588 - Identifyimage failed - ISAM code : 22, object , series $

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.