GE Healthcare · MROptimaBrivo
Cradle will not move in or out nor the Table will move up or down due to home sensor switch misadjustment.
Symptom
Cradle Will Not Move In or Out Table raised normally. LPCA will not drive in or out.
Applies to
- Models
- Optima, Brivo
- Product
- Optima MR360 Alias; HDsv. This concept covers both Optima MR360 and Brivo MR355. Both products are produced in Beijing, China.
- Software
- SV20.1 Applies to Optima MR360 and Brivo MR355, a.k.a. HDsv. Both products are made in Beijing, China.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.