TPS reset fails due to failed PHPS2.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2268865 The long drive cable is not detected at the SRI. Error: 2268865 Host: mr3cov Process: NSP File: SCP:SriManager Line: 3788 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Unknown Coil: Unknown Scan: -1/-1 The long drive cable is not detected at the SRI.

2274508 CFB detected that the table power from the PHPS is low. Bad cable connection to PHPS or bad PHPS is the likely cause. See Long Description Error: 2274508 Host: signa2 Process: NSP File: SCP:CfbHardware Line: 620 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Unknown Coil: Unknown Scan: -1/-1 CFB detected that the table power from the PHPS is low. Bad cable connection to PHPS or bad PHPS is the likely cause.

2268878 The motor controller communication fiber optic loopback test failed. May 16 Mon 09:49:55 2268878 NSP SCP:LongMotionMgr 1892 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Unknown Coil: Unknown Scan: -1/-1 The motor controller communication fiber optic loopback test failed. (SRI to MCPDB)->

2256925 StartupCondition error on SCP resulted in TPS reset failure. See Long Description 

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
Component
5167111 Assembly Patient Handling Power Supply 2 for DVMR 5167111 Assembly, Patient Handling Power Supply 2
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.