Multiple 2267859 GD Power Failure errors due to failed 48 Volt power supply in PDU chassis.

Symptom

Will Not Scan with Gradient Errors.

Error message

2267859 The % axis XGA is reporting a GD power Failure %V undervoltage Example: Mon Nov 29 06:57:58 2010 Error: 2267859 Host: MRDV Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 The Z axis XGA is reporting a GD power failure -17V undervoltage.- Mon Nov 29 06:57:59 2010 Error: 2267859 Host: MRDV Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 The X axis XGA is reporting a GD power failure 5V undervoltage.- Mon Nov 29 06:57:59 2010 Error: 2267859 Host: MRDV Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 The Y axis XGA is reporting a GD power failure 5V undervoltage.-

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5140621-4 48V Power Supply 1000W Condor PN GPMP 900_48 TEAL PN 0870109 5140621-4 48V Power Supply 1000W Condor PN GPMP 900_48 TEAL PN 0870109. Part is being set up for 3.0T DVMR program RSPL. 501139061, 5-Mar-08.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.