2274013 Unused Channel Fault - Short Circuit.

Symptom

RF Bias Unused Channel Fault with HNS coil Cannot scan TL section of HNS Coil, getting RF Bias fault for unused channel 2274013. Other coils scan ok using same A connector.

Error message

2274013 Function RF Bias Unused Channel Fault - Short Circuit A multicoil bias fault was detected see long 2274013 Function RF Bias Unused Channel Fault - Short Circuit A multicoil bias fault was detected Connector: Port A Channels faulted: 0x10 First channel: 5 of 8, counting from 1 Fault value: 1551 mV, Channel: 5-> Extended Error: A short circuit was detected by the RF Hub on a channel that is not known to be connected a coil. This could indicate a coil is plugged in that is not being read by coil ID. Run the Multicoil Bias Diagnostic or Coil Datapath Diagnostic to troubleshoot. Explanation of fields: Connector - The coil connector the fault was detected on. Channels faulted - This is a hexadecimal bit map of all the faulted channels. For example, 0x3 means faults on channels 1 and 2. 0x80000000 means channel 32. First channel - The first channel that faulted, and the number of channels on the connector. Fault value - The voltage seen when the fault was detected, on the channel.

2274029 Unused Channel Fault - Short Circuit A multicoil bias fault was detected while scanning. Reseat all coils and check coil connections. Connector: Port 2 (P2) Channels faulted: 0x8000 First channe

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Component
5250080-3 MCDB 5250080-3 MCDB
Area
MR

Full cause & step-by-step resolution