Scan Aborts with EOS Error due to defective SFP module.

Symptom

Scan Aborts with EOS Error Openspeed system running HFO3 M4. When customer uses the frfse-rpc PSD (intermittently) they will only get half the images reconstructed (only odds). This is because the scan has aborted. Then when they try to download the next series the TPS hangs and the customer has to reset TPS or reboot the system. Operator sees the message "Pulse Sequence Error Occurred". Message log has EOS and other errors logged. See attachments for the review screens and errors logged during these events. MRIhc02494 Openspeed system running HFO3 M4. When customer uses the frfse-rpc PSD (intermittently) they will only get half the images reconstructed (only odds). This is because the scan has aborted. Then when they try to download the next series the TPS hangs and the customer has to reset TPS or reboot the system. Operator sees the message "Pulse Sequence Error Occurred". Message log has EOS and other errors logged. See attachments for the review screens and errors logged during these events. MRIhc02494

Error message

2252713 PSD programming error. Pulse generation was still active when an SSP EOS was received Wed Sep 5 10:09:30 2007 Error: 2252713 Host: EMR1 Process: NSP File: AGP:seq_cntl.c Line: 671 PSD programming error. Pulse generation was still active when an SSP EOS was received.

2218537 Scan stopped, Scan Stopped due to error : PSD programming error. Pulse generation was still active when an SSP EOS was received Sat Nov 1 08:42:22 2008 2218537 1 lx-mr scn MsaScanMgr.m 2852 Scan stopped, Scan Stopped due to error : PSD programming error. Pulse generation was still active when an SSP EOS was received..

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5183329-2 Fiber Optic Transceiver SFP 4Gb Multimode, RoHS-Compliant Note, as of March 24, 2010 this part can not be ordered yet. It will be a FRU. If you cant order the transceiver order the higher FRU.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.