MR-Touch Inline color MR Elastography (MRE) images can not be transferred to PACS due to color MRE images generated by I…
MR-Touch Inline color MR Elastography (MRE) images can not be transferred to PACS due to color MRE images generated by Inline Post-Processing populates values in tags which are invalid per DICOM stand…
Symptom
The MR-Touch Inline color MR Elastography (MRE) images can not be transferred to PACS See Long Description Reference Trackwise PR # 5075441. Problem - The problem reported by the site was that they could not transfer MR Elastography color images to the Fuji Synapse PACS system. Investigation Details – Zhenghui Zhang (PSD Engineer) downloaded site images on engineering system. Michael Figueira (MR Principal Engineer) reviewed the images and observed that color MR Elastography (hereafter referred as MRE) color images were populated as RGB for photometric interpretation tag, which is an Invalid value for MR Images and therefore were rejected by the PACS system.
Applies to
- Models
- Discovery
- Product
- Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- Corrupt Image Database.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID104279. System Software suddenly locked up.