MR-Touch Inline color MR Elastography (MRE) images can not be transferred to PACS due to color MRE images generated by I…

MR-Touch Inline color MR Elastography (MRE) images can not be transferred to PACS due to color MRE images generated by Inline Post-Processing populates values in tags which are invalid per DICOM stand…

Symptom

The MR-Touch Inline color MR Elastography (MRE) images can not be transferred to PACS See Long Description Reference Trackwise PR # 5075441. Problem - The problem reported by the site was that they could not transfer MR Elastography color images to the Fuji Synapse PACS system. Investigation Details – Zhenghui Zhang (PSD Engineer) downloaded site images on engineering system. Michael Figueira (MR Principal Engineer) reviewed the images and observed that color MR Elastography (hereafter referred as MRE) color images were populated as RGB for photometric interpretation tag, which is an Invalid value for MR Images and therefore were rejected by the PACS system.

Applies to

Models
Discovery
Product
Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Area
MR

Full cause & step-by-step resolution