GE Healthcare · MROptima2269307
TPS Reset fails with error 2269307 due to SCP3 board.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2269307 Communication problem between the AGP and SCP boards. 2269307 diagCoord pearlMap.c 2171 Task ID: 0x0 Time: 0 Object: SCP Exception Class: Unknown Function: Software/Firmware/Config: N/A Communication problem between the AGP and SCP boards.
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 2294300-22 MGD SCP3 RoHS Compliant board 2294300-22 MGD SCP3 RoHS Compliant board
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- IApplication Software does not complete the boot process with TPS Download failure Due to Idle Overnight With Power On.
- CWA ID107471. Unable to Download or Reset The TPS MGD or ISE.