GE Healthcare · MROptima22612392269053791162252239504361121430417
TPS reset fails due to missing 24 volts for CAN provided by Driver Module.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2261239 ASC NB IF Board was not detected but is configured. See Long Description Fri Jul 15 11:46:26 2005 2261239 lx-mr NSP SCP:CanAutoConfig 2587 ASC NB IF Board was not detected but is configured in MRconfig.cfg. If this board is present, verify power, RF Amp and CAN communication and perform a TPS Reset. If the SSM is present, set usssType=no in MRconfig.cfg and perform a TPS Reset.-> 2269053 The 1-Wire Network did not detect the isolated 24VDC CAN Power 2269053 Host: GEHCMR02 Process: diagCoord File: diagHw.c Line: 820 Diagnostic: CFB Board Level Diagnostics Iteration: 1 of 1 Action: 1 Task ID: 0xf29a9f0 Time: 79116 Object: SCP Test Task Function: Software/Firmware/Config: N/A The 1-Wire Network did not detect the isolated 24VDC CAN Power. 2252239 Startup Timeout was detected on the CFB. Dec 8 Wed 13:30:36 2252239 NSP SCP:CanAutoConfig 3457 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: cohnoise Coil: HEAD Scan: 50436/1 Startup Timeout was detected on the CFB.-> 2252239 Startup Timeout was detected on the SRI. Dec 8 Wed 13:30:36 2252239 NSP SCP:CanAutoConfig 3457 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communic
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5160446-2 Driver Module NMR DVMR
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- IApplication Software does not complete the boot process with TPS Download failure Due to Idle Overnight With Power On.
- No signal on any Scans With no errors in log Due to Driver Module.