GE Healthcare · MROptima22688168719385376
Cradle loses Home Position due to missing Cradle Latch tension spring.
Symptom
Cradle loses Home Position when driven fully out of scanner. Cradle loses home position and reports "no home found" when brought fully out of the magnet. Customer has to drive it 1.5-2" in before it begins to count.
Error message
2268816 Home switch position appears to have moved. Tue Jun 14 14:51:32 2011 Error: 2268816 Host: SAMCMRIWB Process: NSP File: SCP:LongMotionMonit Line: 1195 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: T1flair Coil: 8HRBRAIN Scan: 1033/21 Home switch position appears to have moved. Press the home button to re-set home position. Last position: -87193 counts New position : -85376 counts
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5184384-2 FRU Kit, DVMR LPCA Cradle Latch See Long Description 5184384-2FRU Kit, DVMR LPCA Cradle Latch Refer to latest Service Documentation.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.