Pular para o conteúdo principal
GE Healthcare · MROptima22688168719385376

Cradle loses Home Position due to missing Cradle Latch tension spring.

Symptom

Cradle loses Home Position when driven fully out of scanner. Cradle loses home position and reports "no home found" when brought fully out of the magnet. Customer has to drive it 1.5-2" in before it begins to count.

Error message

2268816 Home switch position appears to have moved. Tue Jun 14 14:51:32 2011 Error: 2268816 Host: SAMCMRIWB Process: NSP File: SCP:LongMotionMonit Line: 1195 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: T1flair Coil: 8HRBRAIN Scan: 1033/21 Home switch position appears to have moved. Press the home button to re-set home position. Last position: -87193 counts New position : -85376 counts

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5184384-2 FRU Kit, DVMR LPCA Cradle Latch See Long Description 5184384-2FRU Kit, DVMR LPCA Cradle Latch Refer to latest Service Documentation.
Area
MR

Full cause & step-by-step resolution