Pular para o conteúdo principal
GE Healthcare · MRDiscovery226789322677782227556

Fails TPS Reset when booting up to Applications or when Resetting TPS due to Axis XGD Amplifier.

Symptom

Unable to Scan with Safemode Startup of XGD Failed for each Axis See Long Description Unable to scan with errors logged "Safemode Startup of XGD Failed" for every gradient axis seen at same time in error log. See errors, below. Notice that the date/time is the same for each error, and that all axes are reporting the error: X, Y & Z. Fri Mar 27 11:31:06 2009 Error: 2267798 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XGA-SB X axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .-> Fri Mar 27 11:31:06 2009 Error: 2267801 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XGA-SB Y axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .-> Fri Mar 27 11:31:06 2009 Error: 2267804 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XGA-SB Z axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .-> Fri Mar 27 11:31:06 2009 Error: 2267807 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XPS-SB X axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .-> Fri Mar 27 11:31:06 2009 Error: 2267810 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XPS-SB Y axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .-> Fri Mar 27 11:31:06 2009 Error: 2267813 Host: GEMR3T Process: NSP File: SCP:GradientHwo Line: 1907 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Safe mode startup of the XGD (XPS-SB Z axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative .->

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2267893 The %b XPS-SB is reporting an internal VBus overvoltage warning See Long Description 2267893 EM_XGD_PS_VBUS_IN_FAULT The %b XPS-SB is reporting an internal VBus overvoltage warning

2267778 An error on the XGD has prevented TPS reset from completing Successfully Error: 2267778 Host: UCSF-OIMR1 Process: NSPFile: SCP:GradientHwo Line: 1986Exception Class: Unknown Severity: UnknownFunction: Gradient: N/APSD: Unknown Coil: Unknown Scan: -1/-1 An error on the XGD has prevented TPS reset from completing successfully. Please review the log for XGD faults to determine the cause and action(s) to take. Perform another TPS reset and if this error occurs again, contact your service representative.

2227556 *** Reset/Download TPS was not successful. *** 2227556 sdm sdm_main.c 320 *** Reset/Download TPS was not successful. ***

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Component
5148804-2 XGD Power Supply XPS 5148804-2 XGD Power Supply XPS ** 2nd Opinion Part ** Confirm appropriate extended error messages have been read and gradient diagnostics have been executed before placing order for this part. 212002950 May 28, 2009. Promoted this part for DVMR 3.0T NPI program, 212035293, 10 April 2008.