GE Healthcare · MROptima22688512268924
2268851 and 2268924 when booting Applications or when Resetting TPS due to incorrect gradient coil selected.
Symptom
2268851and 2268924 when booting Applications or when Resetting TPS Applications boot successfully but get 2268851 and 2268924 during boot process.
Error message
2268851 Gradient coil ibc inlet temperature sensor %d failed See Long Description Note: "%d" is a variable that will be replaced by the actual failed temperature sensor by the system when the error message is logged. 2268924 Num sensors (config): 4 Num sensors (coil):6 See extended error message. 2268924 Host: MRI Process: NSP File: SCP:SriManagerLine: 5384 Num sensors (config): 4 Num sensors (coil):6-> Extended Error: This error suggests that either the HART ID on the gradient coil was improperly programmed or the system was configured for the wrong gradient coil type. It can safely be ignored, as long as the gradient coil type is properly configured. An incorrectly configured gradient coil type could lead to IBC inlet temperature sensor failure error messages and/or incorrect IBC inlet temperature readings.
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.