GE Healthcare · MROptima226777822569252267863

2267863 The % axis XGA is reporting a I_Coil Wire Fault due to XGA amplifier.

Symptom

Cannot scan with Error: 2267863 The % axis XGA is reporting a I_Coil Wire Fault Cannot scan with Error: 2267863 The % axis XGA is reporting a I_Coil Wire Fault.

Error message

2267778 An error on the XGD has prevented TPS reset from completing Successfully Error: 2267778 Host: UCSF-OIMR1 Process: NSPFile: SCP:GradientHwo Line: 1986Exception Class: Unknown Severity: UnknownFunction: Gradient: N/APSD: Unknown Coil: Unknown Scan: -1/-1 An error on the XGD has prevented TPS reset from completing successfully. Please review the log for XGD faults to determine the cause and action(s) to take. Perform another TPS reset and if this error occurs again, contact your service representative.

2256925 StartupCondition error on XGD StartupCondition resulted in TPS reset failure See Long Description 2256925 SLMAINMR LxMxConsumer services/initialization/ServerInitManager 767 Server Name: SC StartupCondition error on XGD StartupCondition resulted in TPS reset failure.

2267863 The %axis XGA is reporting a I_Coil Wire Fault See Long Description Extended Error: The cables between the coil current sensor and the SB board is loose or has fallen off. Ensure cables J5 and J6 of the XGA are connected to the corresponding axis current sensor at the top of the Power Cabinet.

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
Component
5184566-2 XGD AMPLIFIER 5184566-2 XGD AMPLIFIER
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues