GE Healthcare · MRDiscoveryOptimaBrivo22732162218537
Auto and Manual Prescan Fails Due to bend pins in the RRX module
Symptom
Auto and Manual Prescan Fails
Error message
2273216 Communication Failure.The RF Hub detected that the general and/or R1 banks selected on all boards Fri May 13 14:08:32 2011 Error: 2273216 Host: GMCMR2 Process: NSP File: SCP:RfHubProcess Line: 2346 Exception Class: Unknown Severity: Unknown Function: RF: RF Receive PSD: FRFSE-XL Coil: HDCTL 456 Scan: 619/5 Communication Failure. The RF Hub detected that the general and/or R1 banks selected on all boards was changed unexpectedly during scan to an invalid index. Detected Banks (General, Gain): 0, 142 2218537 Scan stopped, Scan Stopped due to error : RF Hub detected a fault while scanning.. Fri May 13 14:08:34 2011 Error: 2218537 Host: GMCMR2 Process: scn File: ../MsaScanMgr.m Line: 3153 Scan stopped, Scan Stopped due to error : RF Hub detected a fault while scanning..
Applies to
- Models
- Discovery, Optima, Brivo
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5174169 HDv Remote Receiver Module 5174169 HDv Remote Receiver Module
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.