TPS reset fails due to P1 or P2 port was not connected properly to LPCA .

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2256925 StartupCondition error on SCP resulted in TPS reset failure. See Long Description 2256925 MOI_MR LxMxConsumer services/initialization/ServerInitManager 639 Server Name: SC StartupCondition error on SCP resulted in TPS reset failure.

2276804 Coil port connector was disconnected. Scanning will be allowed when the port connector is reconnected. Connector: Port 1 and or 2 P1 P2 See Long Description for extended error. 2276804 Coil port connector was disconnected. Scanning will be allowed when the port connector is reconnected. Connector: Port 1 and or 2 P1 P2. Extended Error: The RF Hub detected that one of the coil ports in the LPCA was disconnected. Specific errors are logged in other messages. Re-seat the port P1 and P2 connector assemblies into the emergency egress connectors inside the LPCA. Check the coil port cable connections at the RF Hub control board and at the LPCA. Run the Coil Datapath Diagnostic that utilizes the MCRv tool to isolate the problem. Possible cause(s): RF Hub Control Board

2276802 Coil port connector(s) have been disconnected. Scanning is inhibited 2276802 ERROR: EM_RF_HUB_QUICK_DISCONNECT_FAULT Coil port connector(s) have been disconnected. Scanni

Applies to

Models
Discovery
Product
Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Area
MR

Full cause & step-by-step resolution