Fails TPS Reset when booting up to Applications or when Resetting TPS Due to Defective RRX Module

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2267503 DTX1 Calibration Data Collection Failure Calibration of the DTX1 failed because the Irf3 could not collect sample data from the 1st RRX module. This could be a communication failure or an analog signal so large it overflowed the RRX See Long Description 2267503 MOI_MR diagCoord dtxCal.cpp 259 Task ID: 0x47da0b0 Time: 40817 Object: AGP Exception Class: Notification Function: Software/Firmware/Config: N/A DTX1 Calibration Data Collection Failure Calibration of the DTX1 failed because the Irf3 could not collect sample data from the 1st RRX module. This could be a communication failure or an analog signal so large it overflowed the RRX.-> Extended Error: DTX calibration uses the Irf3 board to collect sample data from the 1st RRX module. The DTX calibration failed because the data collection from the RRX failed at least one time. This could indicate a problem with the RRX, the Irf3, or the clock and data fiber cables that connect these devices together.

Applies to

Models
Discovery, Optima, Brivo
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Component
5174169 HDv Remote Receiver Module 5174169 HDv Remote Receiver Module
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required