Pular para o conteúdo principal
GE Healthcare · MROptima2267900

XPS Over-current fault due to failed XGD Power Supply.

Symptom

Gradient Fidelity Test failures

Error message

2267900 The %b XPS-SB is reporting an input power Primary Average Over-current fault. 2267900 ERROR: EM_XGD_PS_PRI_AVG_OC The %b XPS-SB is reporting an input power Primary Average Over-current fault. Try increasing scan TR to lower the DC power draw out of the XPS. Extended Error: Ensure proper connection between the XGA and XPS.

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5148804-2 XGD Power Supply XPS 5148804-2 XGD Power Supply XPS ** 2nd Opinion Part ** Confirm appropriate extended error messages have been read and gradient diagnostics have been executed before placing order for this part. 212002950 May 28, 2009. Promoted this part for DVMR 3.0T NPI program, 212035293, 10 April 2008.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.