Pular para o conteúdo principal
GE Healthcare · MROptima226885528131

TPS reset failed with Motion Controller error due to bad cable from PHPS J39 / Pen Panel interface, run M3033 to J3 of M…

TPS reset failed with Motion Controller error due to bad cable from PHPS J39 / Pen Panel interface, run M3033 to J3 of Magnet I/O.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2268855 Communication to the Motion Controller timed out see long Error: 2268855 Host: mr450w Process: diagCoord File: LongMotionMgr Line: 1883 Diagnostic: CFB Board Level Diagnostics Iteration: 1 of 1 Action: 1 Task ID: 0xd10d230 Time: 28131 Object: SCP Function: Patient Handling: Control/Display Communication to the Motion Controller timed out

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5331721 Scan Room Base SHORT Cable FRU see long FRU Kit contains: Run M1302 Run M1303 Run M1304 Run M1310 Run M1311 Run M1312 Run M2025 Run M3030 Run M3031 Run M3300 Run M3301 Run M3302 Run M3303 Run M3304 Run M3306 Run M3307 Run M3308 Run M3309 Run M3310 Run M3311 Run M3312 Run M3313 Run M3315 Run M3316 Run M4005 13W6 Shunt Connector for 16 & 32 Channel Only
Area
MR

Full cause & step-by-step resolution