GE Healthcare · MROptima2267889
2267889 The % axis XPS is reporting a High Voltage I-Bus Over-current Fault due to XGA.
Symptom
Will Not Scan with Gradient Errors.
Error message
2267889 The Z axis XPS is reporting a High Voltage I-Bus Over-current Fault See Long Description Thu Apr 16 08:42:00 2009 Error: 2267889 Host: MR3T Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: sprlio Coil: HEAD Scan: 193/7 The Z axis XPS is reporting a High Voltage I-Bus Over-current Fault.-> Extended Error: Ensure that the XPS and XGA are properly connected, and that the XGA is operating properly by swapping XPS modules if necessary. If fault persists replace the failing XPS
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- 21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
- Component
- 5184566-2 XGD AMPLIFIER 5184566-2 XGD AMPLIFIER
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.