GE Healthcare · MRDiscovery22674251395122674012275705
Fails TPS Reset when booting up to Applications or when Resetting TPS Due to Scan Room Power Supply.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2267425 No Clock Signal to IRF3 board. Error: 2267425 Host: WIMRMR2 Process: diagCoord File: Irf3.cpp Line: 1402 Task ID: 0x493f6b0 Time: 13951 Object: AGP Exception Class: Notification Function: Software/Firmware/Config: N/A No Clock Signal to IRF3 board. The IRF3 board did not detect a clock signal on the J8 fiber optic clock cable. This condition will cause a TPS reset failure.- 2267401 ERROR EM_IRF3_DRVR_DTX_TIMEOUT See Long Description 2267401 ERROR: EM_IRF3_DRVR_DTX_TIMEOUT A DTX module is not sending status information back to the IRF3 board over its dual fiber cable. The command %s could not be completed. DTX Module %d Extended Error: This might be caused by a momentary DVMR link dropout which mighthave been caused by any module connected to the DVMR fiber links. Check that the DTX is functioning and is connected to the IRF3by a dual fiber cable. Run the DTX Board Level Diagnostics. Possible cause(s): NB DTx Module, IRF Board 2275705 The RF Hub clock was not detected during wakeup when expected see long description Error: 2275705 Host: HSSGMR750 Process: NSP File: SCP:RfHubCanHWO Line: 5869 Exception Class: Unknown Severity: Unknown Function: RF: RF Receive PSD: FRFSE-XL
Applies to
- Models
- Discovery
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- 20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
- Component
- 5231801-3 Scan Room Power Supply 16 ch and 32 ch 5231801-3 Scan Room Power Supply, 16 ch and 32 ch
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.