GE Healthcare · MROptima2254212
2254212 SRFD2 RF Amplifier Peak Power Fault due to SRFD2.
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
2254212 SRFD2 RF Amplifier Peak Power Fault. Peak Power from RF Amplifier Exceeds Specification Fri Apr 30 14:23:00 2004 Error: 2254212 Host: lx-mr1 Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2218 SRFD2 RF Amplifier Peak Power Fault. Peak power from RF Amplifier exceeds specification(72.5dBm). Possible causes: the PSD exceeds system limits, RF Gain miscalibration, defective Exciter, or defective RF Amplifier.->
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573 See comments in Resolution Concept for second opinion requirements before ordering this part. SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- IApplication Software does not complete the boot process with TPS Download failure Due to Idle Overnight With Power On.
- CWA ID107471. Unable to Download or Reset The TPS MGD or ISE.