GE Healthcare · MRDiscovery22757092267432
TPs reset fails due to defective SRPS .
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2275709 Communication Failure. The RF Hub failed to recover CAN communication on wakeup. The SCP lost contact with RRX-1 through the fiber link. See Long Description 2275709 NSP SCP:RfHubCanHWO 5951 Exception Class: Unknown Severity: Unknown Function: RF: RF Receive PSD: Probe-P Coil: HNS Head Scan: 2/15 Communication Failure. The RF Hub failed to recover CAN communication on wakeup. The SCP lost contact with RRX-1 through the fiber link. Check RF Hub power, clock status, and CAN cable connections.-> Extended Error: Check RF Hub power from the SRPS and PDB, check the RRX to RF Hub cables, and check the CAN communication cabling. Run the RF Hub Main Board Level Diagnostic, RF Hub Shutdown Diagnostic, and CAN Link Diagnostic 2267432 The Reference Clock to the IRF3 or ERF is not present 2267432 NSP AGP:Irf3RrfHWO 1350 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: FSE-XL Coil: HNS Head Scan: 3/5 The Reference Clock to the IRF3 or ERF is not present.-
Applies to
- Models
- Discovery
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- 20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
- Component
- 5231801-3 Scan Room Power Supply 16 ch and 32 ch 5231801-3 Scan Room Power Supply, 16 ch and 32 ch
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.