Pular para o conteúdo principal
GE Healthcare · MROptima226777822678352256925

gate-drive over-current Errors Due to XGA Amplifier

Symptom

Gate-drive over-current has Occurred on % axis XGA Errors Getting gate-drive over-current has occurred on Z axis XGA error message when trying to scan.

Error message

2267778 An error on the XGD has prevented TPS reset from completing Successfully Error: 2267778 Host: UCSF-OIMR1 Process: NSPFile: SCP:GradientHwo Line: 1986Exception Class: Unknown Severity: UnknownFunction: Gradient: N/APSD: Unknown Coil: Unknown Scan: -1/-1 An error on the XGD has prevented TPS reset from completing successfully. Please review the log for XGD faults to determine the cause and action(s) to take. Perform another TPS reset and if this error occurs again, contact your service representative.

2267835 A gate-drive over-current has occurred on % axis XGA Aug 18 Tue 05:04:17 2267835 NSP SCP:FspMac.cpp 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 A gate-drive over-current has occurred on Z axis XGA. Extended error log: 2267835ERROR: EM_XGD_AS_GD_OC A gate-drive over-current has occurred on %b XGA. If this message persists after TPS resets, the corresponding XGA should be replaced.

2256925 StartupCondition error on XGD StartupCondition resulted in TPS reset failure See Long Description 2256925 SLMAINMR LxMxConsumer services/initialization/ServerInitManager 767 Server Name: SC StartupCondition error on XGD 

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Component
5184566-2 XGD AMPLIFIER 5184566-2 XGD AMPLIFIER
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.