GE Healthcare · CTDiscovery26010009050234
"Scan hardware Stopped Scan" with PDU errors
Symptom
"Scan hardware Stopped Scan" with PDU errors "Scan hardware Stopped Scan" with PDU errorss
Error message
Error: 260100090 The Kss Contactor Status Signal changed state unexpectedly. Contactor Status: Open See long description for details Tue Mar 11 20:26:37 2008 Error: 260100090 Host: tgp Process: tgp subsys File: scIsrService.c Line: 153 Tue Mar 11 20:26:37 2008 Host : Table Gantry Processor Ermes # : 260100090 Exception Class : Abort Severity : Primary File : scIsrService.c Line# : 153 Function : SYSTEM MONITORING : HVDC Monitor Scan Type : Axial Scan: 50234/1/2 PDU KSS Contactor Status. The Kss Contactor Status Signal changed state unexpectedly. Contactor Status: Open
Applies to
- Models
- Discovery
- Product
- Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5324764 GE 14-25A CONTACTOR (Kss) See long description. Part #: 5324764 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: GE 14-25A CONTACTOR FRU: Yes Status: Orderable Parts Offered as: Current: 5324764 Harvested: Not Available Exchange: Not Available Comments: Per Titan RSPL, 212008898, 1/15/2010.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required