GE Healthcare · MROptimaDiscovery2270032226887322678492267600
Fails TPS Reset when booting up to Applications or when Resetting TPS Due to Scan Room Power Supply breaker tripped.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2270032 - VRF detected RRX status issue Mon Sep 21 14:46:30 2009 Error: 2270032 Host: LIBR-MR1 Process: LxMxConsumer File: VrfTrace.cpp Line: 80 Server Name: ADS VRF detected RRX status issue. Check the dual fiber optic cable connection from 2268873 The home and end of travel switches are not detected Mon Sep 21 14:46:30 2009 Error: 2268873 Host: LIBR-MR1 Process: NSP File: SCP:SriManager Line: 3800 Exception Class: Unknown Severity: Unkn 2268873 The home and end of travel switches are not detected. 2267849 The GMC is reporting a clock stop condition 2009 Error: 2267849 Host: LIBR-MR1 Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown The GMC is reporting a clock stop condition Extended Error: Check that the system clock is plugged into the GMC. 2267600 - The IRF3 detected a fault on the fiber links. Status register See Long Description Error: 2267600 Host: LIBR-MR1 Process: NSP File: AGP:Irf3RrfHWO Line: 1418 Exception Class: Unknown Severity: Unknown The IRF3 detected a fault on the fiber links. Status register=0x%X The modules connected to these IRF3 jacks were determined to be the cause of the disruption: %s Extended Error: A momentary pro
Applies to
- Models
- Optima, Discovery
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.