Pular para o conteúdo principal
GE Healthcare · MRDiscovery226772822676002267910

2267910 Power cabinet reporting 48V overvoltage due to PS1 48 V supply in PDU.

Symptom

Gradient Fidelity Diagnostic Error 2267910 The Power Cabinet AUX board is reporting a 48V undervoltage (or overvoltage) See Long Description This error seen when pulsing gradients, running gradient fidelity diagnostic: Fri Apr 10 02:02:51 2009 Error: 2267910 Host: RIC_HDMR2 Process: diagCoord File: FspMac.cpp Line: 320 Diagnostic: Average Forward Power Measurement Circuit Test Iteration: 1 of infinite Action: 1 Task ID: 0x2548ba0 Time: 3996792 Object: SCP Function: Gradient: N/A The Power Cabinet AUX board is reporting a 48V undervoltage (or overvoltage) . Message will state either undervoltage or overvoltage but not both at same time.

Error message

2267728 Unexpected communication loss with XGD. Please TPS reset and try again. If this error occurs again, contact your service representative. See Long Description ERROR: EM_XGD_HWO_COM_FAILURE Extended Error: The SCP unexpectedly lost its Ethernet connection to the XGD. Verify that an Ethernet cable is connected from the GMC to the switch and that the link LED on the GMC Ethernet port is illuminated (top LED). Verify that an Ethernet cable is connected from the SCP to the switch and that the link LED on the SCP Ethernet port is illuminated (top LED). Power down the XGD and verify that the GMC is seated securely in its slot by removing and re-seating it. Power on the XGD and perform a TPS reset. Power down the MGD chassis and verify that SCP is seated securely in its slot by removing and re-seating it. Power on the MGD chassis and perform a TPS reset.

2267600 The IRF3 detected a fault on the fiber links. Status register, J10 - XGD See Long Description 2267600 ERROR: EM_IRF3_FIRST_LINK_DOWN The IRF3 detected a fault on the fiber links. Status register=0x%X The modules connected to these IRF3 jacks were determined to be the cause of the disruption: %s Extended Error: A momentary p

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Component
5140621-4 48V Power Supply 1000W Condor PN GPMP 900_48 TEAL PN 0870109 5140621-4 48V Power Supply 1000W Condor PN GPMP 900_48 TEAL PN 0870109. Part is being set up for 3.0T DVMR program RSPL. 501139061, 5-Mar-08.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.