PDU main breaker rotating handle trips as soon as 48V gradient control breaker is turned on due to defective fluid leak…

PDU main breaker rotating handle trips as soon as 48V gradient control breaker is turned on due to defective fluid leak sensor.Replace leak sensor cable in PGR cabinet wqr.Optima MR450W 21.0 5266125 F…

Symptom

PDU main breaker rotating handle trips as soon as it powers on. Happens only when the 48V gradient control breaker is turned on. PDU main breaker rotating handle trips as soon as it powers on.

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
Component
5266125 Fluid Leak Sensor Kit 5266125 Fluid Leak Sensor Kit
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.