TPS Reset fails with NO XGD MODULE FOUND Error Due to 48V Gradient Control Breaker at the PDU Was Tripped.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2267414 NO XGD MODULE FOUND. The IRF3 board does not see the XGD module. See Long Description Error: 2267414 Host: PHSRADMR02 Process: diagCoordFile: Irf3.cpp Line: 2233Task ID: 0x493f570 Time: 25411 Object: AGPException Class: NotificationFunction: Software/Firmware/Config: N/A NO XGD MODULE FOUND.The IRF3 board does not see the XGD module. The system must have a dual fiber cable between the GMC board (in the Cronus Chassis) of the XGD and the IRF3 board-> Extended Error : The IRF3 did not detect an active fiber cable from the XGD. The XGD must be connected to the IRF3 by a dual fiber cable. The XGD must have a single strand clock fiber cable from the DTX1 module. Verify that the XGD is powered on and that it's clock and data fibers are connected.

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.