GE Healthcare · MROptima2267925
There are greater output tolerances between two types of power supplies in the Chronus chassis.
Symptom
The GMC is reporting a 12V overvoltage. Error 2267925 appears in the system error log while the gradients are pulsing but does not stop the scan. Error will not appear while gradients are at rest.
Error message
2267925 The GMC is reporting a 12V overvoltage Error: 2267925 Host: A131GEMR01 Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: FSE-XL Coil: HNS CTLTOP Scan: 725/2 The GMC is reporting a 12V overvoltage.
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5159513 XGD Cronus cPCI Power Supply 5159513 XGD Cronus cPCI Power Supply 8HP 3U
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.