GE Healthcare · MROptima2273003
Coil id not recognized intermittently due to Coil Cable.
Symptom
Coil ID not recognized intermittently. Only one coil is affected.
Error message
2273003 Coil validation error. An incorrect CRC was detected in the coil ID data from the coil. The coil is programmed incorrectly. See Long Desc. 2273003 Coil validation error. An incorrect CRC was detected in the coil ID data from the coil. The coil is programmed incorrectly. Connector: %b Extended Error: If this happens whenever the coil is plugged in, this is most likely a bad coil. If not, this could indicate a bad connection at the coil interface, a bad coil, bad cables to the RF Hub RFCB, or a bad RFCB. Try connecting another coil. Run the Coil ID Diagnostic or Coil Datapath Diagnostic to troubleshoot cabling issues and to isolate the coil interface from the coil.
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 2416710 SYSTEM CABLE, 1.5T CARDIAC Part #: 2416710Part Type: GP CONSUMABLES Part Weight: 0.00 LBRRepairable: NoReturnable: NoHazardous Material: NDescription: FRU SYSTEM CABLE, 1.5T CARDIACFRU: YesStatus: Orderable
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- IApplication Software does not complete the boot process with TPS Download failure Due to Idle Overnight With Power On.
- CWA ID107471. Unable to Download or Reset The TPS MGD or ISE.