Warning Message The bore temperature sensor is disabled.

Symptom

Warning Message The bore temperature sensor is disabled. Do not scan patients. See long description Customer called InSite this morning as when they arrived they saw a message "The bore temperature sensor is disabled. Do not scan patients." InSite then correctly instructed the operator to NOT scan patients until the problem was rectified.

Error message

2225096 The bore temperature sensor is disabled. Do not scan patients. Error: 2225096 Host: GE-MR Process: NSP Line: 11 File: HPC_PORT_ID:hpc_rd_ipg_cfg The bore temperature sensor is disabled. Do not scan patients. The bore temperature sensor is emulated in the ipg_stage file. Remove emulation flag and Reset TPS to enable monitoring.

Applies to

Models
Ovation
Product
Ovation/MFO
Software
MFO3 Applies to Ovation3 with HP4000 computer running Linux OS.
Area
MR

Full cause & step-by-step resolution