Pular para o conteúdo principal
GE Healthcare · MROptimaDiscovery226890450171

Unable to Scan with RF Air Temperature Faults Due To Customer Coolant Chiller.

Symptom

Unable to Scan with RF Air Temperature Faults

Error message

2268904 RF air temperature warning detected. Future scanning will be inhibited until RF air temperature matches expected. See Long Description Thu Jan 15 14:56:24 2009 Error: 2268904 Host: signa2 Process: NSP File: SCP:SriCanHWO Line: 2030 Exception Class: Unknown Severity: Unknown Function: Patient Handling: Control/Display PSD: Gradient Echo Coil: HEAD Scan: 50171/1 RF air temperature warning detected. Future scanning will be inhibited until RF air temperature matches expected. Temperature difference: 20 0.1 degrees C Tolerance: 20 0.1 degrees C-> Extended Error The Heat Exchanger (HEC) manages RF coil air temperature. This message indicates that the HEC was unable to achieve the requested RF air temperature. This may be caused by facility water temperature or flow not as required.

Applies to

Models
Optima, Discovery
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues