Startup tests fails constantly after PSD check.
Symptom
Startup tests fails constantly after PSD check. Clutch test step is never reached. Gantry motions are not possible Startup tests fails constantly after PSD check. Clutch test step is never reached. Gantry motions are not possible. Recurrent error message is Error ID: 1434 Description: prepareMoveToLimitCommand failed .
Error message
MOTION_INTERRUPTED_BY_HW_RCU Before the error message: RCU was not released : code = 0x0 14:58:12 41579 [mIntHandler] RCUControl::handleKeyChangeInterrupt()- RCUKeysRegister_ = 0x4001 14:58:12 41579 [mIntHandler] AxesSynchronizer::stop() - completionReason = MOTION_INTERRUPTED_BY_HW_RCU 14:58:12 41579 [mIntHandler] SafetyControl::handleInterrupts(): received RCU Key int. 14:58:13 41819 [mIntHandler] RCUControl::handleKeyChangeInterrupt()- RCUKeysRegister_ = 0x4041 14:58:13 41819 [mIntHandler] AxesSynchronizer::stop() - completionReason = MOTION_INTERRUPTED_BY_HW_RCU 14:58:13 41819 [mIntHandler] SafetyControl::handleInterrupts(): received RCU Key int. 14:58:13 41899 [mIntHandler] RCUControl::handleKeyChangeInterrupt()- RCUKeysRegister_ = 0x4040 Error # 1434 Description: prepareMoveToLimitCommand failed 14:58:19 Task:POPUP Activity:POPUP_MESSAGE Parameters:ServerType=DialogPopupServer|title=Message|text=Startup tests failed. Error 1434 - General error, means one of the motion condition tests have failed. however, this can be one of many causes (RCU stuck, constant Enable signal, MCM is down...).
Applies to
- Models
- Discovery
- Product
- Discovery NM/CT 670 Discovery NM/CT 670 DSV NM/CT 670 DSV670 Discovery NM 630 Discovery NM 630
- Component
- 5394765 RCU Discovery NM/CT Spare Part 5394765 RCU Discovery NM/CT Spare Part
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- VG Fanbeam collimators copying cor correction values
- Hawkeye X-Ray patient images ring artifacts
- Ventri failed system startup test, "LAST AXIS MOVEMENT FAILED"
- VG detectors radial 1 or radial 2 will not radial out in fast mode with RCU
- Ventri CAN Network Maping Initialization - ERROR_GCU_CAN_NETWORK_INITIALIZATION
- Venti or Discovery 530/570 Internal Host, DIB and GCU IP Address