GE Healthcare · MROptimaBrivo2254791
Unable to scan with Power Monitor Faults.
Symptom
Unable to scan with Power Monitor Faults
Error message
2254791 The Universal Power Monitor X detected a Pulse Width fault on the Narrow Band Body Amplifier Sun Nov 20 12:43:16 2005 Error: 2254791 Host: HDMR Process: NSP File: SCP:UPM Line: 1231 The Universal Power Monitor 1 detected a Pulse Width fault on the Narrow Band Body Amplifier. Detected Width = 130 uS. Max Width Limit = 0 uS.
Applies to
- Models
- Optima, Brivo
- Product
- Optima MR360 Alias; HDsv. This concept covers both Optima MR360 and Brivo MR355. Both products are produced in Beijing, China.
- Software
- SV20.1 Applies to Optima MR360 and Brivo MR355, a.k.a. HDsv. Both products are made in Beijing, China.
- Area
- MR
Full cause & step-by-step resolution