GE Healthcare · MRDiscoveryOptimaBrivo2270016227003122675863388122674312267597
RRX Module is not communicating with VRF(icn) or IRF3 due to defective RRX module.
Symptom
Cannot reset TPS with IRF3 not ready message Cannot reset TPS with IRF3 not ready message.
Error message
2270016 ADS1 VRF did not receive correct device ID from RRX module in scan room. RRX does not exist or is not connected to VRF. The VRF is located at ICN1(ICN2), Fri Feb 27 07:06:54 2009 Error: 2270016 Host: mr1 heri Process: LxMxConsumer File: VrfTrace.cpp Line: 80 Server Name: ADS1 VRF did not receive correct device ID from RRX module in scan room. RRX does not exist or is not connected to VRF. The VRF is located at ICN1(ICN2), if ADS(ADS1) reports this error. VRF ID: 0 Device ID: 2- 2270031 RRX to VRF fiber link ERROR. Check the dual fiber optic cable connection from the RRX to the VRF. Nov 12 Fri 16:13:07 2270031 LxMxConsumer VrfTrace.cpp 80 Server Name: ADS1 RRX to VRF fiber link ERROR. Check the dual fiber optic cable connection from the RRX to the VRF. Check power to RRX module. The VRF is located at ICN1(ICN2), if ADS(ADS1) reports this error. VRF Register: 0x30 Descriptions: 0x131000000000-> 2267586 The RF Hub reports that the following RRX do not have an active fiber signal See in the long description. Tue Apr 13 10:45:30 2010 Error: 2267586 Host: nmr5 Process: diagCoord File: rfHub.cpp Line: 787 Task ID: 0x47d8d10 Time: 33881 Object: AGP Exception Class: Unknown Functi
Applies to
- Models
- Discovery, Optima, Brivo
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 5174169 HDv Remote Receiver Module 5174169 HDv Remote Receiver Module
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- Corrupt Image Database.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID104279. System Software suddenly locked up.