Auto prescan fails due to P1 or P2 port was not connected properly to LPCA .

Symptom

Auto and Manual Prescan Fails

Error message

2276804 Coil port connector was disconnected. Scanning will be allowed when the port connector is reconnected. Connector: Port 1 and or 2 P1 P2 See Long Description for extended error. 2276804 Coil port connector was disconnected. Scanning will be allowed when the port connector is reconnected. Connector: Port 1 and or 2 P1 P2. Extended Error: The RF Hub detected that one of the coil ports in the LPCA was disconnected. Specific errors are logged in other messages. Re-seat the port P1 and P2 connector assemblies into the emergency egress connectors inside the LPCA. Check the coil port cable connections at the RF Hub control board and at the LPCA. Run the Coil Datapath Diagnostic that utilizes the MCRv tool to isolate the problem. Possible cause(s): RF Hub Control Board

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
Area
MR