Environmental Locks activate due to restricted airflow for GOC.

Symptom

Environmental Locks activate when starting a new scan Environmental Locks activate when starting each scan. Operator is prompted to bypass the Environmental Locks before a new scan can be started.

Error message

2269034 The operator room temperature has exceeded the upper warning limits. Please make the necessary adjustments to prevent equipment damage. Wed Dec 29 09:17:09 2010 Error: 2269034 Host: DVMR Process: envMon File: monitor.c Line: 503 Exception Class: Abort Function: Global Operator Console: Environmental Sensors/Controls The operator room temperature has exceeded the maximum upper limits. Please make the necessary adjustments to prevent equipment damage. Instance: 0 Value instance: 0 Value: 35.31 Upper Limit: 35.00

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Area
MR