Pular para o conteúdo principal
GE Healthcare · MRDiscovery22678882267904

Will Not Scan with Gradient Errors.

Symptom

Will Not Scan with Gradient Errors.

Error message

2267888 The %b XPS is reporting a Low Voltage I-Bus Over-current Fault.-> Extended Error: Ensure that the XPS and XGA are properly connected, and that the XGA is operating properly by swapping XPS modules if necessary. If fault persist replace the failing XPS. Note: In the actual error message seen on the scanner, "%b" will be replaced with either the X, Y, or Z axis.

2267904 The %b XPS-SB is reporting a VBus Unbalanced Warning. See Long Description 2267904 Host: MR3T Process: NSP File: SCP:FspMac.cpp Line: 320 Exception Class: Unknown Severity: Unknown ERROR: EM_XGD_PS_VOLT_UNBALANCE The %b XPS-SB is reporting a VBus Unbalanced Warning.

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Area
MR