Guided Install Hardware Tab had the Spectroscopy RF Amplifier selected with No Spectroscopy RF Amplifier.Fails TPS Reset…

Guided Install Hardware Tab had the Spectroscopy RF Amplifier selected with No Spectroscopy RF Amplifier.Fails TPS Reset when booting up to Applications or when Resetting TPS.In Guided Install Hardwar…

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2261240 ASC BB IF Board was not detected but is configured in MRconfig.cfg. 1121427986 3 1 Fri Jul 15 11:46:26 2005 2261240 2 lx-mr NSP SCP:CanAutoConfig 2597 ASC BB IF Board was not detected but is configured in MRconfig.cfg. If this board is present, verify power, RF Amp and CAN Communication and perform a TPS Reset. If the SSM is present, set usssType=no in MRconfig.cfg and perform a TPS Reset.-> EN 351 SR 162

2252239 Startup Timeout was detected on the RF BB Amp Interface. 1121427986 2 1 Fri Jul 15 11:46:26 2005 2252239 2 lx-mr NSP SCP:CanAutoConfig 2576 Startup Timeout was detected on the RF BB Amp Interface.-> EN 162 SR 351

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Area
MR