Fails TPS Reset when booting up to Applications or when Resetting TPS Due to Scan Room Power Supply breaker tripped.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2269020 A Scan Room Power Supply Board 1-Wire Device, srps_fru_exc_pio, was not discovered on the 1-Wire Network Extended Error: This can happen if the device was not powered or properly connected when the system was booted. Verify the 1-Wire Device is properly connected. See system block diagrams if necessary. TPS Reset may be necessary to clear this error.

2269019 The Reference Clock Board Identification 1-Wire Device, clkref_bid, was not discovered on the 1-Wire Network Extended Error: This can happen if the device was not powered or properly connected when the system was booted. Verify the 1-Wire Device is properly connected. See system block diagrams if necessary. TPS Reset may be necessary to clear this error.

2269018 An Exciter Board 1-Wire Device, nbexchdv_bid, was not discovered on the 1-Wire Network Extended Error: This can happen if the device was not powered or properly connected when the system was booted. Verify the 1-Wire Device is properly connected. See system block diagrams if necessary. TPS Reset may be necessary to clear this error.

2269029 A XGD GMC Board 1-Wire Device, xgd_gmc_bid, was not discovered on the 1-Wire Network Extended Error: This can happen if 

Applies to

Models
Discovery
Product
Discovery MR450 Also called DVMR. This product is compatible with 1.5T 20.1, 20.1_IB1, IB2, 21.0 and DV22.0 software systems only.
Software
21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
Area
MR

Full cause & step-by-step resolution