Pular para o conteúdo principal
GE Healthcare · MRDiscovery226911618944

2269116 The maximum number of SRPS fault recovery attempts has been reached Due to SRPS Scan Room Power Supply.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2269116 The maximum number of SRPS fault recovery attempts has been reached and the SRPS is still indicating errors via the 1-Wire Network. A manual SRPS recovery may be necessary see long Error: 2269116 Host: MR02 Process: diagCoord File: oneWire.c Line: 1834 Task ID: 0x6db6aa0 Time: 18944 Object: AGP Exception Class: Notification Function: Software/Firmware/Config: N/A The maximum number of SRPS fault recovery attempts has been reached and the SRPS is still indicating errors via the 1-Wire Network. A manual SRPS recovery may be necessary

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Component
5231801-3 Scan Room Power Supply 16 ch and 32 ch 5231801-3 Scan Room Power Supply, 16 ch and 32 ch
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.