Pular para o conteúdo principal
GE Healthcare · MRDiscovery225213322185372252756

Scanning inhibited with logged error codes 2252133, 2252756, 2218537 due to a software timing problem on release 20.1 so…

Scanning inhibited with logged error codes 2252133, 2252756, 2218537 due to a software timing problem on release 20.1 software.

Symptom

Scanning inhibited with logged error codes 2252133, 2252756, 2218537 See Long Description Scanning inhibited and these messages seen in error log: Thu Oct 21 16:54:34 2010 Error: 2252133 Host: DVMR450 Process: NSP File: SCP:sdo_segment_upload Line: 1925 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: FSE-XL Coil: HNSp CTL234 Scan: 2032/16 SDO Communication Failure was detected on the CANopen Protocol. Node Detecting Failure = SCP. Node ID = 11. Index = 0x3002. SubIndex = 0x0. Error Code = 0x1001.-> Thu Oct 21 16:54:38 2010 Error: 2252756 Host: DVMR450 Process: NSP File: AGP:agp_drv.c Line: 1692 Exception Class: Unknown Severity: Unknown Function: Acquisition/Pulse Generation: Pulse Generation PSD: FSE-XL Coil: HNSp CTL234 Scan: 2032/16 The bore temperature sequence was not updated by the SCP within the 5 second interval. This indicates that the SCP is not monitoring the bore temperature during scan. Possible SCP functionality failure. Thu Oct 21 16:54:40 2010 Error: 2218537 Host: DVMR450 Process: scn File: ../MsaScanMgr.m Line: 3168 Scan stopped, Scan Stopped due to error : The bore temperature sequence was not updated by the SCP within the 5 second interval. This indicates that the SCP is not monitoring the bore temperature during scan. Possible SCP functionality failure.

Error message

2252133 SDO Communication Failure was detected on the CANopen Protocol Node Detecting Failure = SCP Wed Aug 25 08:04:47 2004 2252133 2 MR-OC0 NSP SCP:sdo_segment_upload 1408 SDO Communication Failure was detected on the CANopen Protocol. Node Detecting Failure = SCP. Node ID = 4. Index = 0x3003. SubIndex = 0x0. Error Code = 0x1002.->

2218537 Scan stopped: The bore temperature sequence was not updated by the SCP. See long description Tue Jan 13 18:04:32 2009 2218537 mr3t scn MsaScanMgr.m 3038 Arrêt acquisition. Scan Stopped due to error : The bore temperature sequence was not updated by the SCP within the 5 second interval. This indicates that the SCP is not monitoring the bore temperature during scan. Possible SCP functionality failure..

2252756 The bore temperature sequence was not updated by the SCP within the 5 second interval. This indicates that the SCP is not monitoring the bore temperature during scan. Possible SCP functionality failure. See Long Description Error: 2252756 Host: DVMR450 Process: NSP File: AGP:agp_drv.c Line: 1692 Exception Class: Unknown Severity: Unknown Function: Acquisition/Pulse Generation: Pulse Generation PSD: FSE-XL Coil: HNSp CTL234 Scan: 2032/16 T

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.