2267857 Gradient Distortion fault and 2268220 Gradient Fidelity and Static Diagnostics failure has occurred on the % axi…

2267857 Gradient Distortion fault and 2268220 Gradient Fidelity and Static Diagnostics failure has occurred on the % axis due to defective XGD amplifier.

Symptom

Gradient Diagnostic Failure

Error message

2268220 Gradient Static Diagnostic failed limit test in register Nov 10 Wed 10:13:35 2268220 diagCoord DiagStatic.cpp 546 Diagnostic: Gradient Static Iteration: 1 of 1 Action: 1 Task ID: 0xdafb150 Time: 306175 Object: SCP Function: Software/Firmware/Config: N/A GRADIENT DIAGNOSTIC - Gradient Static Diagnostic failed limit test in register: AC/IERR_INT.X. Value = 2740da5fH. Limits = fff38000H, 000c8000H. Refer to Service Level message for action.

2267857 A gradient distortion fault has occurred on the X axis.->

Applies to

Models
Optima
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
Component
5184566-2 XGD AMPLIFIER 5184566-2 XGD AMPLIFIER
Area
MR

Full cause & step-by-step resolution