GE Healthcare · MROptima2267728
XGD Not Found Due to 48 Volt Power Supply(s).Application Software does not complete the boot process with TPS Download f…
XGD Not Found Due to 48 Volt Power Supply(s).Application Software does not complete the boot process with TPS Download failure.Replace 48 volt power supply(s) in PDU-wqr.Optima MR450W.2267728 ERROR
Symptom
Application Software does not complete the boot process with TPS Download failure
Error message
2267728 Unexpected communication loss with XGD. Please TPS reset and try again. If this error occurs again, contact your service representative. See Long Description ERROR: EM_XGD_HWO_COM_FAILURE Extended Error: The SCP unexpectedly lost its Ethernet connection to the XGD. Verify that an Ethernet cable is connected from the GMC to the switch and that the link LED on the GMC Ethernet port is illuminated (top LED). Verify that an Ethernet cable is connected from the SCP to the switch and that the link LED on the SCP Ethernet port is illuminated (top LED). Power down the XGD and verify that the GMC is seated securely in its slot by removing and re-seating it. Power on the XGD and perform a TPS reset. Power down the MGD chassis and verify that SCP is seated securely in its slot by removing and re-seating it. Power on the MGD chassis and perform a TPS reset.
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- 21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
- Component
- 5140621-4 48V Power Supply 1000W Condor PN GPMP 900_48 TEAL PN 0870109 5140621-4 48V Power Supply 1000W Condor PN GPMP 900_48 TEAL PN 0870109. Part is being set up for 3.0T DVMR program RSPL. 501139061, 5-Mar-08.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- IApplication Software does not complete the boot process with TPS Download failure Due to Idle Overnight With Power On.
- CWA ID107471. Unable to Download or Reset The TPS MGD or ISE.