GE Healthcare · MROvation2220025
Intermittent nuisance message TCU error 2220025.
Ovation MFO TCU error 2220025.
Symptom
Intermittent nuisance message TCU error 2220025. OPENMR4.292a_M3_0415 software running on MFO4. System could scan but the error was appearing in the error log.
Error message
2220025 TCU channel #1 temperature is 33.4 C, more than 0.7 C below the prescribed setpoint 40.0 C. Mon May 23 15:39:57 2005 Error: 2220025 Host: RMIMROVATION Process: port_mg r File: PMTCU.cc Line: 65 TCU channel #1 temperature is 33.4 C, more than 0.7 C below the prescri bed setpoint 40.0 C.
Applies to
- Models
- Ovation
- Product
- Ovation with EXCITE
- Software
- OPENMR4-MFO4 Applies to Ovation4 (Excite) only. Ovation4 (MFO4) is 11.x based software.
- Area
- MR