CWA Unable to Graphically Prescribe off of a 3-plane Localizer, the Graphic Prescription Window Does not Come Up.
Symptom
Unable to Graphically Prescribe off of a 3-plane Localizer, the Graphic Prescription Window Does not Come Up One of the windows / panes is missing for 3 plane-loc. See long description. The tech scans a 3 plane localizes and then wants to graphically prescribe off of that, but the graphic prescription window does not come up, an error comes up: "no valid localizer exists for current prescription".
Error message
2219864 An internal error occured in List/Select function at 3PLANE GRx
Applies to
- Models
- Optima
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- 21.0 Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750). Only for Optima MR450W (wide bore) systems. (Not to be used on Discovery MR450 or Discovery MR750).
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA ID100513. Will Not Scan with Gradient Errors.
- CWA ID103384. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID103582. Unable To Scan With Driver Module Errors.
- CWA ID104077. Computer Will Not Boot with the message on the screen that says Do Not Interact With System.
- CWA ID104279. System Software suddenly locked up.
- IApplication Software does not complete the boot process with TPS Download failure Due to Idle Overnight With Power On.